ProductModel: | BR-SSS-A |
Classification: | IEC 61215-2-2021 Series |
RetailPrice: | 0 |
★ Product launched: 2011 ★ Earliest user: BJSPI ★ Other users: Ingeer Certification, LERRI, Guizhou Quality Institution, FJMI-National Photovoltaic Industry Metrology and Testing Center....
BR-SSS-A
Class AAA Steady-State Solar Simulator &
Multi-Sections Variable Speed I-V Sweep System, Micro-Inverter…
Perform standards:IEC 61215-2:2021, IEC 61730-2:2020 (draft), IEC 60904-1-2:2019,
IEC 61853-1:2011, IEC 61853-2:2016, UL 1703-2018, IEC 60904.11, IEC 60891:2009...
Test Item |
Classifications |
Irr. W/m2 |
Temp. °C |
Time or Dose |
Initial Stabilization |
AAA/CCC |
800~1000 |
50 |
5/10 KW·h/m2 |
Final Stabilization |
CCC |
800~1000 |
50 |
— |
Hot-spot endurance test |
AAA/BBB |
700/800~1300 |
50 |
1 or 5h |
Temperature test |
— |
1000 |
40 |
— |
Measure of temperature coefficients |
BBB |
1000 |
20→55 |
Continuous |
Performance at low irradiance |
BBA |
200 |
25 |
— |
Performance at STC and NMOT |
BBA |
1000 / 800 |
25 / 20 |
— |
Maximum power determination |
AAA |
1000 |
25 |
— |
Arc test (for UL1703 method B) |
— |
800 |
20~30 |
15min |
HTC power rating (IEC 61853-1) |
BBB |
1000 |
75 |
— |
■ Light source system
□ Spectral Level: Class AAA
Spectral match c lass A calibration range: 300~1200nm
□ Effective Irradiation area: 2400×1400mm (W×D) or more
□ Irradiance range: 700~1300W/m 2 or 980~1020W/m 2
□ Irradiance resolution: 1W/m2
□ Testing temperature range: 25~60°C±2 °C or wider
□ Irradiance measurement: by standard cell (I-V sweep) or radiation meter (exposure)
□ Irradiance measure: by standard cell for I-V measurt and pyranometer for exposure
□ I-V measurement: Isc, Voc, Pm, Imp, Vmp, FF, η, Rs, Rp
□ Automatically generating and storing I-V and P-V curve
□ Xenon lamps matrix is arranged, optimization lighting fixtures angles adjustable.
□ Filters: Class A of AM1.5 for Crystal Si-Solar or Amorphous Si-solar (Si-α-Si)
■ Electronic load and I-V measurement system
□ Constant speed I-V sweep: 10~1000ms
Constant speed sweep: 10~20ms
Multi-sections variable speed I-V sweep: 100~1000s or settings
for high capacitance & high efficiency module I-V measure
□ Single module maximum test power: 750W (500V/30A or 120V/60A)
□ Voltage measurement accuracy: ±(0.025%+0.05%FS)
□ Current measurement accuracy: ±(0.05%+0.05%FS)
□ Power measurement accuracy: ±(0.2%+0.2%FS)
□ Display accuracy: 1mV / 0.1mA
□ Temperature measurement precision: ±0.5°C
□ Irradiance measurement: by standard cell
□ Three modes of I-V measurement:
Single channel single measurement for maximum power determination
Single channel repeat measurement for temperature coefficients
Option: Multichannel cycle measurement for outdoor array
■ Micro-inverter for maximum power point tracking (MPPT)
□ Voltage and power: 22~60V, 60~130V, 130~250V, 600W*1/300W*2
□For stabilization, outdoor exposure, NMOT, temperature test...
■ Automatic Non-Uniformity Measurement System
□ Mechanical transmission system program control, along the X-Y axis
horizontal - longitudinal reciprocating scanning, data acquisition and processing.
□ Scanning time can be set.
□ Measurement point more than 64 (IEC 60904-10).
□ Automatic calculate average irradiance and nonuniformity.
■ Hot-Spot Endurance Test
□ Class AAA for determination of cell selection and shading area determination,
and Class BBB for exposure.
■ Shading plate for hot-spot endurance test:
□ Crystalline silicon module
Cells shaded plates: 50%,20%,20%,10%,5%
Can satisfy the 0 ~ 100% within 5% increasing or decreasing.
□ Thin-film module
1~5 pieces cell shaded plates horizontal transmission,
Ordinal sequence covering 1~5 pcs cells, according to current find Imp×99%, determine shaded plates number.
■ Initial Stabilization (Pre-exposure test / LID) :
Irradiation 5/10/60KW·h/m² or software setting
■ Software Interface :